대역폭 | 350, 500 MHz, 1, 2 GHz |
---|---|
샘플링 속도 | 10 Gsample/s |
메모리 | 25 Mpts |
채널 수 | 2 or 4 채널 |
디스플레이 | 10.4 인치 컬러 디스플레이 |
MSO5034B DPO5034B |
MSO5054B DPO5054B |
MSO5104B DPO5104B |
MSO5204B DPO5204B |
||||||
---|---|---|---|---|---|---|---|---|---|
Input Channels | 4 | ||||||||
Bandwidth | 350 MHz | 500 MHz | 1 GHz | 2 GHz | |||||
Rise Time (Calculated) | 1 ns | 700 ps | 350 ps | 175 ps | |||||
DC Gain Accuracy | ±1.5%, derated at 0.10%/°C above 30 °C | ||||||||
<td class="cellrowb | |||||||||
Effective Number of Bits (Typical) | 6 bits (10 divisionp-p sine wave input at instrument bandwidth, 100 mV/div, 50 Ω Input Impedance, maximum sample rate, 1 k point record length) | ||||||||
Random Noise (RMS, typical, sample mode, full BW, 50 Ω) | |||||||||
1 mV/div | 173 μV | 178 μV | 68 μV | 70 μV | |||||
2 mV/div | 216 μV | 236 μV | 128 μV | 158 μV | |||||
5 mV/div | 231 μV | 281 μV | 214 μV | 307 μV | |||||
10 mV/div | 305 μV | 340 μV | 315 μV | 485 μV | |||||
20 mV/div | 504 μV | 523 μV | 547 μV | 791 μV | |||||
50 mV/div | 1.15 mV | 1.17 mV | 1.29 mV | 1.82 mV | |||||
100 mV/div | 2.40 mV | 2.46 mV | 3.08 mV | 4.75 mV | |||||
1 V/div | 22.96 mV | 22.98 mV | 23.15 mV | 29.58 mV | |||||
Maximum Sample Rate (All channels) | 5 GS/s | 5 GS/s | 5 GS/s | 5 GS/s | |||||
Maximum Sample Rate (1 or 2 channels) | 5 GS/s | 5 GS/s | 10 GS/s | 10 GS/s | |||||
Maximum Equivalent Time Sampling Rate | 400 GS/s | ||||||||
Maximum Record Length with Standard Configuration | 25 M | 25 M (4 ch) 50 M (1 or 2 ch) |
|||||||
Maximum Record Length with Option 5RL | 50 M | 50 M (4 ch) 125 M (1 or 2 ch) |
|||||||
Maximum Record Length with Option 10RL | 125 M | 125 M (4 ch) 250 M (1 or 2 ch) |
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
MSO5034B DPO5034B |
MSO5054B DPO5054B |
MSO5104B DPO5104B |
MSO5204B DPO5204B |
||
---|---|---|---|---|---|
Input Channels | 4 | ||||
Bandwidth | 350 MHz | 500 MHz | 1 GHz | 2 GHz | |
Rise Time (Calculated) | 1 ns | 700 ps | 350 ps | 175 ps | |
DC Gain Accuracy | ±1.5%, derated at 0.10%/°C above 30 °C | ||||
±40 V
30 Vp-p≤200 MHz
10 Vp-p>200 MHz
25 ms
±75 ns
500 MS/s (2 ns resolution)
25 M Standard
Up to 40 M with Record Length options
16.5 GS/s (60.6 ps resolution)
10k points centered around the trigger
1 ns
200 ps
500 MHz at minimum input swing; higher toggle rates can be achieved at higher amplitudes
DC, AC, HF Rej (attenuates >50 kHz), LF Rej (attenuates <50 kHz), Noise Reject (reduces sensitivity)
External (auxiliary input) 1 MΩ
External (auxiliary input)
Provided as part of Opt. SR-AUTO. Trigger on Start of Frame, Type of Frame (Data, Remote, Error, or Overload), Identifier, Data, Identifier and Data, EOF, Missing Ack, Bit Stuff Error, and CRC Error on CAN buses up to 1 Mb/s
Provided as part of Opt. SR-AUTO. Trigger on Sync, Identifier, Data, Identifier and Data, Wakeup Frame, Sleep Frame, and Error on LIN buses up to 1 Mb/s
Provided as part of Opt. SR-AUTO. Trigger on Indicator Bits (Normal, Payload, Null, Sync, Startup), Cycle Count, Header Fields (Indicator Bits, Identifier, Payload Length, Header CRC, and Cycle Count), Identifier, Data, Identifier and Data, End Of Frame, and Error on FlexRay buses up to 10 Mb/s
Provided as part of Opt. SR-AERO. Trigger on Sync, Command Word, Status Word, Data Word, Idle Time, and Error on MIL-STD-1553 buses up to 1 Mb/s
Provided as part of Opt. SR-ENET. Trigger on Start of Packet, MAC Ad
350 MHz, 5 GS/s, 25 M record length, 4-channel digital phosphor oscilloscope
500 MHz, 5 GS/s, 25 M record length, 4-channel digital phosphor oscilloscope
1 GHz, 10/5 GS/s (2/4 ch), 50 M/25 M record length, 4-channel digital phosphor oscilloscope
2 GHz, 10/5 GS/s (2/4 ch), 50 M/25 M record length, 4-channel digital phosphor oscilloscope
350 MHz, 5 GS/s, 25 M record length, 4+16 channel mixed signal oscilloscope
500 MHz, 5 GS/s, 25 M record length, 4+16 channel mixed signal oscilloscope
1 GHz, 10/5 GS/s (2/4 ch), 50 M/25 M record length, 4+16 channel mixed signal oscilloscope
2 GHz, 10/5 GS/s (2/4 ch), 50 M/25 M record length, 4+16 channel mixed signal oscilloscope
These bundled items must be purchased at the same time as the instrument purchase.